Diagnostic performance of reduced-dose CT with a hybrid iterative reconstruction algorithm for the detection of hypervascular liver lesions: a phantom study.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 27957640.
- Also identified by DOI 10.1007/s00330-016-4687-6.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To investigate the diagnostic performance of reduced-dose CT with a hybrid iterative reconstruction (IR) algorithm for the detection of hypervascular liver lesions. Thirty liver phantoms with or without simulated hypervascular lesions were scanned with a 320-slice CT scanner with control-dose (40 mAs) and reduced-dose (30 and 20 mAs) settings. Control-dose images were reconstructed with filtered back projection (FBP), and reduced-dose images were reconstructed with FBP and a hybrid IR algorithm. Objective image noise and the lesion to liver contrast-to-noise ratio (CNR) were evaluated quantitatively. Images were interpreted independently by 2 blinded radiologists, and jackknife alternative free-response receiver-operating characteristic (JAFROC) analysis was performed. Hybrid IR images with reduced-dose settings (both 30 and 20 mAs) yielded significantly lower objective image noise and higher CNR than control-dose FBP images (P < .05). However, hybrid IR images with reduced-dose settings had lower JAFROC1 figure of merit than control-dose FBP images, although only the difference between 20 mAs images and control-dose FBP images was significant for both readers (P < .01). An aggressive reduction of the radiation dose would impair the detectability of hypervascular liver lesions, although objective image noise and CNR would be preserved by a hybrid IR algorithm. • A half-dose scan with a hybrid iterative reconstruction preserves objective image quality. • A hybrid iterative reconstruction algorithm does not improve diagnostic performance. • An aggressive dose reduction would impair the detectability of low-contrast lesions.
Medical subject headings
- Algorithms
- Liver Neoplasms
- Phantoms, Imaging
- Radiation Dosage
- Radiographic Image Interpretation, Computer-Assisted
- Tomography, X-Ray Computed