Radiation dose reduction using 100-kVp and a sinogram-affirmed iterative reconstruction algorithm in adolescent head CT: Impact on grey-white matter contrast and image noise.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 27966043.
- Also identified by DOI 10.1007/s00330-016-4679-6.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To retrospectively evaluate the image quality and radiation dose of 100-kVp scans with sinogram-affirmed iterative reconstruction (IR) for unenhanced head CT in adolescents. Sixty-nine patients aged 12-17 years underwent head CT under 120- (n = 34) or 100-kVp (n = 35) protocols. The 120-kVp images were reconstructed with filtered back-projection (FBP), 100-kVp images with FBP (100-kVp-F) and sinogram-affirmed IR (100-kVp-S). We compared the effective dose (ED), grey-white matter (GM-WM) contrast, image noise, and contrast-to-noise ratio (CNR) between protocols in supratentorial (ST) and posterior fossa (PS). We also assessed GM-WM contrast, image noise, sharpness, artifacts, and overall image quality on a four-point scale. ED was 46% lower with 100- than 120-kVp (p < 0.001). GM-WM contrast was higher, and image noise was lower, on 100-kVp-S than 120-kVp at ST (p < 0.001). CNR of 100-kVp-S was higher than of 120-kVp (p < 0.001). GM-WM contrast of 100-kVp-S was subjectively rated as better than of 120-kVp (p < 0.001). There were no significant differences in the other criteria between 100-kVp-S and 120-kVp (p = 0.072-0.966). The 100-kVp with sinogram-affirmed IR facilitated dramatic radiation reduction and better GM-WM contrast without increasing image noise in adolescent head CT. • 100-kVp head CT provides 46% radiation dose reduction compared with 120-kVp. • 100-kVp scanning improves subjective and objective GM-WM contrast. • Sinogram-affirmed IR decreases head CT image noise, especially in supratentorial region. • 100-kVp protocol with sinogram-affirmed IR is suited for adolescent head CT.
Medical subject headings
- Gray Matter
- Radiation Dosage
- Radiographic Image Interpretation, Computer-Assisted
- Tomography, X-Ray Computed
- White Matter