A Systematic Investigation of Parameters Influencing Droplet Rain in the Listeria monocytogenes prfA Assay - Reduction of Ambiguous Results in ddPCR.
Where this comes from
- Record sourced from PubMed, PMID 27992475.
- Also identified by DOI 10.1371/journal.pone.0168179 and PMC identifier 5167268.
- Licence recorded as CC BY.
- The licence permits redistribution, so the abstract is shown in full and the full text is available from the publisher.
Abstract
The droplet digital polymerase chain reaction (ddPCR) determines DNA amounts based upon the pattern of positive and negative droplets, according to Poisson distribution, without the use of external standards. However, division into positive and negative droplets is often not clear because a part of the droplets has intermediate fluorescence values, appearing as "rain" in the plot. Despite the droplet rain, absolute quantification with ddPCR is possible, as shown previously for the prfA assay in quantifying Listeria monocytogenes. Nevertheless, reducing the rain, and thus ambiguous results, promotes the accuracy and credibility of ddPCR. In this study, we extensively investigated chemical and physical parameters for optimizing the prfA assay for ddPCR. While differences in the concentration of all chemicals and the dye, quencher and supplier of the probe did not alter the droplet pattern, changes in the PCR cycling program, such as prolonged times and increased cycle numbers, improved the assay.
Medical subject headings
- Bacterial Proteins
- Listeria monocytogenes
- Peptide Termination Factors
- Polymerase Chain Reaction