Incidence of Ceramic Liner Malseating After Ceramic-on-Ceramic Total Hip Arthroplasty Associated With Osteolysis: A 5- to 15-Year Follow-Up Study.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 28012723.
- Also identified by DOI 10.1016/j.arth.2016.11.044.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The aim of our study was to evaluate the clinical and radiographic outcomes of malseating of the acetabular liner in ceramic-on-ceramic total hip arthroplasty (THA). Outcomes for 160 ceramic-on-ceramic THAs, contributed by 116 women and 39 men, were evaluated. Clinical and radiographic measurements were obtained over a 5- to 15-year follow-up for analysis. Liner malseating was identified in 20% of cases. Outcomes for 32 cases with liner malseating (group A) were compared to outcomes for 128 joints with correct liner seating (group B). The Harris hip score at the last follow-up was 90.1 for group A and 89.6 for group B. Osteolysis was identified in 5 cases in group A (15.6%), compared to 3 cases in group B (P < .001). No significant between-group differences were identified with regard to ceramic fracture, audible squeaking, loosening of components, and revision THA. The mean annual liner wear rate was comparable between groups, 0.0045 mm/y for group A and 0.0039 mm/y for group B. The 10-year Kaplan-Meier survivorship, based on an end point of revision THA, was 100% for group A and 99.0% for group B. Over a moderate-length follow-up of 5-15 years, malseating of the acetabular liner was not associated with negative clinical outcomes or THA survivorship. Malseating did increase the incidence of osteolysis, a risk factor for adverse effects. Long-term follow-up studies are needed to fully quantify the effects of malseating of the acetabular liner.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Ceramics
- Hip Joint
- Osteolysis
- Prosthesis Failure
Anatomy
- hip