A biomimetic mesh for treating female stress urinary incontinence.

Lai, Kuilin; Zhang, Jiazhen; Wang, Guoshuai; Luo, Xin; Liu, Man; Zhang, Xinzhi; Yuan, Yuyu; Liu, Bin et al. · Biofabrication · 2017

basic_science · Level V

Where this comes from

Abstract

For patients with medium to severe incontinence, sub-urethral support surgery has a high cure rate, but using synthetic meshes leads to some complications such as mesh erosion/exposure and thigh pain. Autologous or acellular extracellular matrix grafts present few complications but have a high recurrence rate. Regensling<sup>™</sup> is a new sling product made of a synthetic material with a biomimetic structure, aiming to provide long-term mechanical support with a lower complication rate. To assess the safety and effectiveness of Regensling<sup>™</sup>, both in vitro and in vivo experiments were performed. The mesh was implanted in the subcutaneous, intramuscular and sub-urethral regions of rabbits. At 4, 12, and 26 weeks post-implantation, the animals were executed and the implants were studied for their mechanical and biocompatible properties. Compared to the control material, the Regensling<sup>™</sup> was covered by a thin layer of fibrous tissue with good compliance, and had a milder inflammatory response. During the period of the experiment, Regensling<sup>™</sup> showed stable strength with an increasing trend over time.

Medical subject headings