Clinical Evidence Supporting US Food and Drug Administration Premarket Approval of High-Risk Otolaryngologic Devices, 2000-2014.

Rathi, Vinay K; Wang, Bo; Ross, Joseph S; Downing, Nicholas S; Kesselheim, Aaron S; Gray, Stacey T · Otolaryngol Head Neck Surg · 2017

cross_sectional · Level IV

Where this comes from

Abstract

The US Food and Drug Administration (FDA) approves high-risk medical devices based on premarket pivotal clinical studies demonstrating reasonable assurance of safety and effectiveness and may require postapproval studies (PAS) to further inform benefit-risk assessment. We conducted a cross-sectional analysis using publicly available FDA documents to characterize industry-sponsored pivotal studies and PAS of high-risk devices used in the treatment of otolaryngologic diseases. Between 2000 and 2014, the FDA approved 23 high-risk otolaryngologic devices based on 28 pivotal studies. Median enrollment was 118 patients (interquartile range, 67-181), and median duration of longest primary effectiveness end point follow-up was 26 weeks (interquartile range, 16-96). Fewer than half were randomized (n = 13, 46%), blinded (n = 12, 43%), or controlled (n = 10, 36%). The FDA required 23 PASs for 16 devices (70%): almost two-thirds (n = 15, 65%) monitored long-term performance, and roughly one-third (n = 8, 35%) focused on subgroups. Otolaryngologists should be aware of limitations in the strength of premarket evidence when considering the use of newly approved devices.

Medical subject headings