The value of serial metal ion levels in following up patients with metal-on-metal hip arthroplasty.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 28106230.
- Also identified by DOI 10.5301/hipint.5000432.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The sensitivity of cobalt (Co) and chromium (Cr) ion-levels in detecting poorly performing metal-on-metal hip implants is low. This study proposes that serial changes in ion-levels are a more accurate marker of arthroplasties at risk. Serial metal ion-levels and implant data of 285 patients with ASR resurfacing or replacement were studied. Patient and implant characteristics were analysed using univariate and multivariate analyses. 111 (39%) had revision surgery. Time since index surgery (p<0.001), acetabular inclination (p<0.001), their interaction (p<0.001) and femoral head size (p = 0.01) were significant variables. Head size (≤51 mm) had lower Co and Cr levels (p = 0.01). Cr/Co showed marginal decrease over time in the unrevised group and no decrease prerevision. Repeated measurement of ion-levels were higher in the revision group suggesting that serial measurements rather than absolute values may have a role to play in predicting implant failure.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Metal-on-Metal Joint Prostheses
- Osteoarthritis, Hip
- Prosthesis Failure