In vivo tribological performance of 231 metal-on-metal hip articulations.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 28124364.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
231 cobalt-chromium metal-on-metal (METASUL design) retrieved articulations (297 retrieved components) were examined for their in vivo wear behaviour. A running-in period of about one year is observed where the in vivo wear rate for the whole articulation is approximately 35 m/year. After this running-in period, the measured in vivo wear rate of the whole articulation decreases to about 5 m/year. The metal-on-metal articulation can be considered as a very low wear articulation system. The combination of the specially designed components is mandatory to assure this low wear behaviour. In case of a mismatched bearing (stainless steel head - cobalt-chromium inlay), large amount of wear was measured. (Hip International 2002; 2: 73-6).
Anatomy
- hip