Repeatability and agreement of ocular biometry measurements: Aladdin versus Lenstar.
cross_sectional · Level IV
Where this comes from
- Record sourced from PubMed, PMID 28130351.
- Also identified by DOI 10.1136/bjophthalmol-2016-309365.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To assess the repeatability and agreement between the Aladdin and Lenstar biometers. Three consecutive measurements of the right eye of 102 subjects were acquired by the Aladdin (V.1.1.3) and the Lenstar (V.1.0.3), respectively, with a random order of which biometer to use first. Parameters compared included axial length (AL), anterior chamber depth (ACD), flat keratometry (Kf), steep keratometry (Ks), mean keratometry (Km), J<sub>0</sub>, J<sub>45</sub> and white-to-white (WTW) distance. Intraocular lens (IOL) power formulas included SRK/T (Sanders-Retzlaff-Kraff/Theoretical), Holladay 1, Hoffer Q and Haigis. Repeatability (S<sub>r</sub>), repeatability limit (r) and Bland-Altman limits of agreement (LoA) were calculated. S<sub>r</sub> was marginally better with the Lenstar for AL (0.02 mm), ACD (0.02 mm) and WTW distance (0.07 mm), but S<sub>r</sub> was marginally better with the Aladdin for Kf (0.10 dioptre (D)), J<sub>0</sub> (0.07 D) and J<sub>45</sub> (0.06 D). In terms of agreement, the mean difference was small and non-statistically significant between the two devices for ACD, Ks and J<sub>45</sub>. The mean difference for Kf, Km, J<sub>0</sub> and WTW distance were all statistically significantly lower with the Aladdin than the Lenstar for the exception of WTW distance. The Bland-Altman LoA displayed narrow ranges implying good agreement with the exception for WTW distance. For IOL formulas, the mean difference was very small and not statistically significantly different. The LoA were narrow (within 0.50 D), implying acceptable use of each device interchangeably. High levels of repeatability and agreement were found between the Aladdin and Lenstar suggesting that the devices may be used interchangeably in normal eyes with refractive error.
Medical subject headings
- Anterior Eye Segment
- Axial Length, Eye
- Biometry
- Diagnostic Techniques, Ophthalmological