Isolation and concentration of bacteria from blood using microfluidic membraneless dialysis and dielectrophoresis.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 28276545.
- Also identified by DOI 10.1039/c6lc01277a and PMC identifier 5894507.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
A microfluidic system that combines membraneless microfluidic dialysis and dielectrophoresis to achieve label-free isolation and concentration of bacteria from whole blood is presented. Target bacteria and undesired blood cells are discriminated on the basis of their differential susceptibility to permeabilizing agents that alter the dielectrophoretic behavior of blood cells but not bacteria. The combined membraneless microdialysis and dielectrophoresis system isolated 79 ± 3% of Escherichia coli and 78 ± 2% of Staphylococcus aureus spiked into whole blood at a processing rate of 0.6 mL h<sup>-1</sup>. Collection efficiency was independent of the number of target bacteria up to 10<sup>5</sup> cells. Quantitative PCR analysis revealed that bacterial 16S rDNA levels were enriched more than 307-fold over human DNA in the fraction recovered from the isolation system compared with the original specimen. These data demonstrate feasibility for an instrument to accelerate the detection and analysis of bacteria in blood by first isolating and concentrating them in a microchamber.
Medical subject headings
- Bacteremia
- Bacteria
- Bacterial Load
- Electrophoresis
- Microfluidic Analytical Techniques