Dosimetric impact of contouring and needle reconstruction uncertainties in US-, CT- and MRI-based high-dose-rate prostate brachytherapy treatment planning.

Rylander, Susanne; Buus, Simon; Pedersen, Erik M; Bentzen, Lise; Tanderup, Kari · Radiother Oncol · 2017

case_series · Level IV

Where this comes from

Abstract

The purpose was to evaluate the dosimetric impact of target contouring and needle reconstruction uncertainties in an US-, CT- and MRI-based HDR prostate BT treatment planning. US, CT, and MR images were acquired post-needle insertion in 22 HDR-BT procedures for 11 consecutive patients. Dose plans were simulated for an US-, CT- and MRI-based HDR-BT treatment planning procedure. Planning uncertainties in US- and CT-based plans were evaluated using MRI-based planning as reference. Target (CTV<sub>Prostate</sub>) was re-contoured on MRI. Dose results were expressed in total equivalent dose given in 2Gy fractionation dose for EBRT (46Gy) plus 2 HDR-BT fractions. Uncertainties in US- and CT-based planning caused the planned CTV<sub>Prostate</sub>-D<sub>90%</sub> to decrease with a mean of 2.9±5.0Gy (p=0.03) and 2.9±2.9Gy (p=0.001), respectively. The intra-observer contouring variation on MRI resulted in a mean variation of 1.6±1.5Gy in CTV<sub>Prostate</sub>-D<sub>90%</sub>. Reconstruction uncertainties on US resulted in a dose variation of±3Gy to the urethra, whereas data for CT were not available for this. Uncertainties related to contouring and reconstruction in US- and CT-based HDR-BT treatment plans resulted in a systematic overestimation of the prescribed target dose. Inter-modality uncertainties (US and CT versus MR) were larger than MR intra-observer uncertainties.