Follow-Up Study on Fetal CT Radiation Dose in Japan: Validating the Decrease in Radiation Dose.
cross_sectional · Level IV
Where this comes from
- Record sourced from PubMed, PMID 28328259.
- Also identified by DOI 10.2214/AJR.16.17316.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
In 2011, we collected data on fetal CT radiation dose to determine the diagnostic reference level (DRL); however, continuous evaluation of the DRL is necessary. The hypothesis of this study is that the fetal CT radiation dose has decreased, and we predict a widespread use of iterative reconstruction (IR). We also predict that the national decrease in exposure is because of the DRL reported as a result of the previous national study. Various testing protocols from each site were summarized as part of the study results. The minimum, one-fourth (25th percentile), median, three-fourths (75th percentile), and maximum values were obtained for volume CT dose index (CTDI<sub>vol</sub>), dose-length product (DLP), and scan length of 120 fetal CT examinations. The trends for IR usage and tube voltage were also investigated. Compared to the results of the 2011 study (n = 119), the minimum, 25th percentile, median, and 75th percentile values for CTDI<sub>vol</sub> and DLP have decreased for the tabulated results in 2015 (n = 120). The 75th percentile value for CTDI<sub>vol</sub> was 4.9 mGy, which is 43% of the previous value. IR was used in 70% of the sites. The radiation dose was significantly lower among groups that used IR. Four years passed between our initial survey on DRL and the present follow-up survey, and it appears that the previous report sufficiently fulfilled its objective and role in contributing to the decrease in DRL observed in this follow-up study.
Medical subject headings
- Prenatal Diagnosis
- Radiation Dosage
- Radiation Exposure
- Radiation Protection
- Tomography, X-Ray Computed