VMAT plus a few computer-optimized non-coplanar IMRT beams (VMAT+) tested for liver SBRT.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 28341061.
- Also identified by DOI 10.1016/j.radonc.2017.02.018.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To propose a novel treatment approach, designated VMAT+, involving addition of <5 IMRT beams with computer-optimized non-coplanar orientations to VMAT, and evaluate it for liver Stereotactic Body Radiation Therapy (SBRT). VMAT+ is investigated as an alternative for (1) coplanar VMAT and (2) multi-beam non-coplanar treatment. For fifteen patients with liver metastases, VMAT+ plans were compared with (1) dual-arc VMAT and (2) 25-beam, non-coplanar treatment with computer-optimized beam orientations (25-NCP). All plans were generated fully automatically for delivery of the highest feasible tumor Biologically Effective Dose (BED). OAR doses, intermediate-dose-spillage, dose-compactness, and measured delivery times were evaluated. With VMAT+ the maximum achievable tumor BED was equal to that of 25-NCP. Conversely, VMAT resulted in a lower tumor BED in 5 patients. Compared to VMAT, VMAT+ yielded significant dose reductions in OARs. Intermediate-dose-spillage and dose-compactness were significantly improved by 9.8% and 17.3% (p≤0.002), respectively. Treatment times with VMAT+ were only enhanced by 4.1min on average, compared to VMAT (8.4min). Improvements in OAR sparing with 25-NCP, compared to VMAT+, were generally modest and/or statistically insignificant, while delivery times were on average 20.5min longer. For liver SBRT, VMAT+ is equivalent to time-consuming treatment with 25 non-coplanar beams in terms of achievable tumor BED. Compared to VMAT, OAR sparing and intermediate-dose-spillage are significantly improved, with minor increase in delivery time.
Medical subject headings
- Liver Neoplasms
- Radiosurgery
- Radiotherapy Planning, Computer-Assisted
- Radiotherapy, Intensity-Modulated