Prevalence of Unanticipated Events Associated With MRI Examinations: A Benchmark for MRI Quality, Safety, and Patient Experience.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 28356198.
- Also identified by DOI 10.1016/j.jacr.2017.01.043.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To determine the prevalence of unanticipated events (UE) associated with MRI examinations in a multicenter academic radiology department. UE reported by MRI technologists for examinations performed between June 2013 and November 2014 on 17 scanners in a university- and community-affiliated single health system were retrospectively reviewed. Events were categorized into the following categories: (1) problems with orders and scheduling; (2) scan delays; (3) unanticipated foreign bodies; (4) non-contrast-related patient events (eg, patient motion, discomfort, claustrophobia, need for sedation); (5) contrast-related patient events; and (6) technical acquisition issues. Logistic regression analysis of UE was performed after adjusting based on scanner site (university- vs community-affiliated hospital), hospital setting (outpatient [OP] vs mixed outpatient and inpatient [OP/IP]), and timing of the scan (business days vs weekends/holidays). A total of 34,587 MRI examinations were assessed (87% university-affiliated; 58% OP) with 5775 (16.7%) UE (1.9% of patients had more than one category event). Rates of UE for each category were as follows: 1.9% orders and scheduling, 3.3% delays in scan, 0.5% foreign bodies, 10.4% non-contrast-related patient events, 1.3% contrast-related patient events, and 1.5% technical issues. After adjustment for location of scanner, clinical setting, and timing of the scan, the rate of overall UE was significantly higher in university-affiliated sites (coefficient, 0.09 [95% confidence interval (CI), 0.07-0.10]; P < .001), in scans performed in the mixed OP/IP setting (coefficient, 0.09 [95% CI, 0.08-0.09]; P < .001), and in scans performed during weekends/holidays (coefficient, 0.02 [95% CI, 0.01-0.04]; P < .001). UE associated with MRI examinations are common (16.7%), with the majority being patient-related issues unrelated to contrast administration.
Medical subject headings
- Appointments and Schedules
- Benchmarking
- Magnetic Resonance Imaging
- Medical Order Entry Systems