Intraoperative Bioprosthetic Valve Dysfunction Causing Severe Mitral Regurgitation.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 28359486.
- Also identified by DOI 10.1016/j.athoracsur.2016.09.019.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Structural degeneration of bioprosthetic valves usually occurs gradually over time. Failure of a bioprosthetic valve immediately after implantation is extremely rare. Possible causes include obstruction of valve leaflets from preserved subvalvular tissue during chordal-sparing mitral valve replacement (MVR) or strut entrapment by suture loops. We report 2 cases of acute bioprosthetic mitral valve (MV) dysfunction involving newly implanted Perimount Theon (Edwards Lifesciences, Irvine, CA) bioprostheses, causing severe transvalvular mitral regurgitation (MR). Rapid diagnosis was achieved with the use of intraoperative transesophageal echocardiography (TEE). Operative assessment and examination of the explanted valves could not determine a definite cause of failure in either case.
Medical subject headings
- Bioprosthesis
- Heart Valve Prosthesis Implantation
- Intraoperative Complications
- Mitral Valve Insufficiency
- Prosthesis Failure