The comparability of HR-pQCT bone measurements is improved by scanning anatomically standardized regions.
cross_sectional · Level IV
Where this comes from
- Record sourced from PubMed, PMID 28391447.
- Also identified by DOI 10.1007/s00198-017-4010-7 and PMC identifier 5526099.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We investigated the sensitivity of distal bone density, structure, and strength measurements by high-resolution peripheral quantitative computed tomography (HR-pQCT) to variability in limb length. Our results demonstrate that HR-pQCT should be performed at a standard %-of-total-limb-length to avoid substantial measurement bias in population study comparisons and the evaluation of individual skeletal status in a clinical context. High-resolution peripheral quantitative computed tomography (HR-pQCT) measures of bone do not account for anatomic variability in bone length: a 1-cm volume is acquired at a fixed offset from an anatomic landmark. Our goal was to evaluate HR-pQCT measurement variability introduced by imaging fixed vs. proportional volumes and to propose a standard protocol for relative anatomic positioning. Double-length (2-cm) scans were acquired in 30 adults. We compared measurements from 1-cm sub-volumes located at the default fixed offset, and the average %-of-length offset. The average position corresponded to 4.0% ± 1.1 mm for radius, and 7.2% ± 2.2 mm for tibia. We calculated the RMS difference in bone parameters and T-scores to determine the measurement variability related to differences in limb length. We used anthropometric ratios to estimate the mean limb length for published HR-pQCT reference data, and then calculated mean %-of-length offsets. Variability between fixed vs. relative scan positions was highest in the radius, and for cortical bone in general (RMS difference Ct.Th = 19.5%), while individuals had T-score differentials as high as +3.0 SD (radius Ct.BMD). We estimated that average scan position for published HR-pQCT reference data corresponded to 4.0% at the radius, and 7.3% at tibia. Variability in limb length introduces significant bias to HR-pQCT measures, confounding cross-sectional analyses and limiting the clinical application for individual assessment of skeletal status. We propose to standardize scan positioning using 4.0 and 7.3% of total bone length for the distal radius and tibia, respectively.
Medical subject headings
- Bone Density
- Radius
- Tibia
- Tomography, X-Ray Computed