Increasing strength of a biomedical Ti-Nb-Ta-Zr alloy by alloying with Fe, Si and O.

Stráský, Josef; Harcuba, Petr; Václavová, Kristína; Horváth, Klaudia; Landa, Michal; Srba, Ondřej; Janeček, Miloš · J Mech Behav Biomed Mater · 2017

basic_science · Level V

Where this comes from

Abstract

Low-modulus biomedical beta titanium alloys often suffer from low strength which limits their use as load-bearing orthopaedic implants. In this study, twelve different Ti-Nb-Zr-Ta based alloys alloyed with Fe, Si and O additions were prepared by arc melting and hot forging. The lowest elastic modulus (65GPa) was achieved in the benchmark TNTZ alloy consisting only of pure β phase with low stability due to the 'proximity' to the β to α'' martensitic transformation. Alloying by Fe and O significantly increased elastic modulus, which correlates with the electrons per atom ratio (e/a). Sufficient amount of Fe/O leads to increased yield stress, increased elongation to fracture and also to work hardening during deformation. A 20% increase in strength and a 20% decrease in the elastic modulus when compared to the common Ti-6Al-4V alloy was achieved in TNTZ-Fe-Si-O alloys, which proved to be suitable for biomedical use due to their favorable mechanical properties.

Medical subject headings