Pulse Transit Time Measurement Using Seismocardiogram, Photoplethysmogram, and Acoustic Recordings: Evaluation and Comparison.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 28436909.
- Also identified by DOI 10.1109/JBHI.2017.2696703.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This work proposes a novel method of pulse transit time (PTT) measurement. The proximal arterial location data are collected from seismocardiogram (SCG) recordings by placing a micro-electromechanical accelerometer on the chest wall. The distal arterial location data are recorded using an acoustic sensor placed inside the ear. The performance of distal location recordings is evaluated by comparing SCG-acoustic and SCG-photoplethysmogram (PPG) measurements. PPG and acoustic performances under motion noise are also compared. Experimental results suggest comparable performances for the acoustic-based and PPG-based devices. The feasibility of each PTT measurement method is validated for blood pressure evaluations and its limitations are analyzed.
Medical subject headings
- Photoplethysmography
- Pulse Wave Analysis
- Signal Processing, Computer-Assisted