Serum Metal Levels for Diagnosis of Adverse Local Tissue Reactions Secondary to Corrosion in Metal-on-Polyethylene Total Hip Arthroplasty.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 28511949.
- Also identified by DOI 10.1016/j.arth.2017.04.016.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Recently, corrosion at the head-neck junction in metal-on-polyethylene bearing surface total hip arthroplasty (THA) has been recognized as a cause of adverse local tissue reactions (ALTRs). Serum metal levels have been advocated as a tool for the diagnosis of ALTR; however, no prior studies have specifically examined their utility. The purpose of this study was to determine the optimal cutoff values for serum cobalt and chromium levels in diagnosing ALTR after metal-on-polyethylene bearing surface THA. We reviewed 447 consecutive patients with serum metal levels tested at our institution and identified 64 patients with a metal-on-polyethylene bearing who had axial imaging or underwent reoperation to confirm the presence or absence of ALTR. Receiver-operating characteristic curves were produced to identify cutoff thresholds to optimize sensitivity, and diagnostic test performance was characterized. Forty-four of the 64 patients (69%) were positive for an ALTR. The best test for the diagnosis of ALTR was the serum cobalt level (area under the curve [AUC] = 99%). A threshold cutoff of ≥1.0 ng/mL had a sensitivity of 100%, specificity of 90%, positive predictive value (PPV) of 96%, and negative predictive value (NPV) of 100%. Serum chromium levels were also diagnostic (AUC = 87%). A threshold cutoff of ≥0.15 ng/mL had a sensitivity of 100%, specificity of 50%, PPV of 81%, and NPV of 100%. Finally, serum cobalt-to-chromium ratio was also helpful for diagnosis (AUC = 90%). A threshold cutoff value of 1.4 for the cobalt-to-chromium ratio offered a sensitivity of 93%, specificity of 70%, PPV of 87%, and NPV of 82%. Measurement of serum cobalt level with a threshold value of 1.0 ng/mL in our experience is the best test for identifying the presence of ALTR in patients with a metal-on-polyethylene THA. Measurement of chromium level and the ratio of cobalt-to-chromium levels are also of value.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Chromium
- Cobalt
- Hip Prosthesis
- Metals
- Polyethylene
- Prosthesis Failure
Anatomy
- hip