Verifying tumor position during stereotactic body radiation therapy delivery using (limited-arc) cone beam computed tomography imaging.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 28532606.
- Also identified by DOI 10.1016/j.radonc.2017.04.022.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Proof of tumor position during stereotactic body radiotherapy (SBRT) delivery is desirable. We investigated if cone-beam CT (CBCT) scans reconstructed from (collimated) fluoroscopic kV images acquired during irradiation could show the dominant tumor position. Full-arc CBCT scans were reconstructed using FDK filtered back projection from 38kV fluoroscopy datasets (16 patients) continuously acquired during volumetric modulated spine SBRT. CBCT-CT match values were compared to the average spine offset values found using template matching+triangulation of the individual kV images. Multiple limited-arc CBCTs were reconstructed from fluoroscopic images acquired during lung SBRT of an anthropomorphic thorax phantom using 20-180° arc lengths and for 3 breath-hold lung SBRT patients. Differences between 3D CBCT-CT match results and average spine offsets found using template matching+triangulation were 0.1±0.1mm for all directions (range: 0.0-0.5mm). For limited-arc CBCTs of the thorax phantom, the automatic 3D CBCT-CT match results for arc lengths of 80-180° were ≤1mm. 20° CBCT reconstruction still allowed for positional verification in 2D. (Limited-arc) CBCT reconstructions of kV images acquired during irradiation can identify the dominant position of the tumor during treatment delivery.
Medical subject headings
- Cone-Beam Computed Tomography
- Lung Neoplasms
- Radiosurgery
- Radiotherapy, Image-Guided