Long-term Results of Scandinavian Total Ankle Replacement.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 28535724.
- Also identified by DOI 10.1177/1071100717698695.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The purpose of this study was to evaluate the long-term results of the Scandinavian Total Ankle Replacement (STAR) implant in a single center. Between 1997 and 2002, 34 consecutive ankles were replaced with the STAR ankle prosthesis and reviewed in 2016. Patients were followed both radiologically and clinically; clinical outcomes were measured with the Kofoed ankle score and subjective measurements. The median follow-up period was 159 months (13.3 years; range, 13-202 months). The implant survival rate was 93.9% (95% confidence interval [CI], 92.2%-86.7%) at 5 years, 86.7% (95% CI, 69.3%-96.2%) at 10 years, and 63.6% (95% CI, 30.8%-89.0%) at 15 years. A total of 15 ankles (44%) had been revised: 2 by fusion, 3 by further replacement, 5 by insert exchange, and 5 by bone grafting for marked osteolysis. The improvements in the Kofoed ankle score and Kofoed ankle score for pain were statistically significant at every postoperative evaluation point compared to the preoperative values (all P < .0001), and subjective patient satisfaction was high. Our results were satisfactory and in agreement with previously reported findings of STAR implants. Level IV, retrospective case series.
Medical subject headings
- Ankle Joint
- Arthroplasty, Replacement, Ankle
- Joint Prosthesis
- Osteoarthritis
- Prosthesis Design
Anatomy
- ankle