Comparative Incidence of Patellofemoral Complications Between 2 Total Knee Arthroplasty Systems in a Multicenter, Prospective Clinical Study.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 28600108.
- Also identified by DOI 10.1016/j.arth.2017.04.014.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Cumulative incidence rates (CIRs) of patellofemoral complications (PCCs) and patellofemoral symptomatic crepitus (SC) using a new knee system-total knee arthroplasty (NEW-TKA) were compared with those of a currently available product-total knee arthroplasty (CA-TKA). Twenty-two investigators prospectively enrolled 704 patients into a study using CA-TKA; 364 received a posterior-stabilized configuration. Twenty-three investigators (19 from the CA-TKA study) enrolled 1138 patients with NEW-TKA; 584 received a posterior-stabilized configuration. CIRs were estimated with Kaplan-Meier methods. CA-TKA had 32 PCCs (6.15% CIR at 1 year and 8.26% at 2 years). NEW-TKA had 19 PCCs (3.15% CIR at 1 year and 4.11% at 2 years). CA-TKA had 15 SCs, 12 before 2 years (CIR = 3.67%); NEW-TKA had 5 SCs before 2 years (CIR = 1.21%). NEW-TKA had a statistically lower CIR of overall PCC and SC (log-rank P values = .018 for PCC and .017 for SC). Interim 1- and 2-year CIRs of PCC and SC in NEW-TKA were less than half of those for CA-TKA, which is a promising trend for this new implant.
Medical subject headings
- Arthroplasty, Replacement, Knee
- Postoperative Complications
Anatomy
- femur
- knee