Total Hip Intraoperative Femur Fracture: Do the Design Enhancements of a Second-Generation Tapered-Wedge Stem Reduce the Incidence?

Colacchio, Nicholas D; Robbins, Claire E; Aghazadeh, Mehran S; Talmo, Carl T; Bono, James V · J Arthroplasty · 2017

retrospective_cohort · Level III

Where this comes from

Abstract

Intraoperative femur fracture (IFF) is a well-known complication in primary uncemented total hip arthroplasty (THA). Variations in implant instrumentation design and operative technique may influence the risk of IFF. This study investigates IFF between a standard uncemented tapered-wedge femoral stem and its second-generation successor with the following design changes: size-specific medial curvature, proportional incremental stem growth, modest reduction in stem length, and distal lateral relief. A single experienced surgeon's patient database was retrospectively queried for IFF occurring during primary uncemented THA using a standard tapered-wedge femoral stem system or a second-generation stem. All procedures were performed using soft tissue preserving anatomic capsule repair and posterior approach. The primary outcome measure was IFF. A z-test of proportions was performed to determine significant difference between the 2 stems with respect to IFF. Patient demographics, Dorr classification, and implant characteristics were also examined. Forty-one of 1510 patients (2.72%) who received a standard tapered-wedge femoral stem sustained an IFF, whereas 5 of 800 patients (0.63%) using the second-generation stem incurred an IFF. No other significant associations were found. A standard tapered-wedge femoral stem instrumentation system resulted in greater than 4 times higher incidence of IFF than its second-generation successor used for primary uncemented THA. Identifying risk factors for IFF is necessary to facilitate implant system improvements and thus maximize patient outcomes.

Medical subject headings

Anatomy