Intraoperative clinical testing overestimates the therapeutic window of the permanent DBS electrode in the subthalamic nucleus.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 28699067.
- Also identified by DOI 10.1007/s00701-017-3255-4.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Intraoperative test stimulation is established to optimize target localization in STN DBS, but requires a time-consuming awake surgery in off-medication state. The aim of this study was to compare the thresholds of stimulation-induced effects of test stimulation and the permanent electrode. Fifty-nine PD patients receiving bilateral STN DBS were clinically examined with stepwise increasing monopolar stimulation during surgery and DBS programming at matched stimulation depths. Thresholds of therapeutic and side effects were obtained from standardized examination protocols. Postoperative stimulation via the permanent electrode caused side effects at a significantly lower threshold than predicted during intraoperative test stimulation (P < 0.001); whereas sufficient therapeutic effects were achieved at significantly higher thresholds (P < 0.001). Intraoperative testing may lead to an overestimation of the therapeutic window. The two different electrodes lead to distinct spreading of the electric field in the STN and surrounding tissues that causes different volume of tissue activated (VTA). Clinicians involved in DBS surgery and programming should be aware of the differences in both stimulation settings, concerning electrodes geometry, stimulation modes as well as the impact of time. Therapeutic and side effects of permanent stimulation are not predictable by intraoperative test stimulation. Test stimulation may be an orientating test for very low thresholds of side effects instead.
Medical subject headings
- Deep Brain Stimulation
- Electrodes, Implanted
- Intraoperative Neurophysiological Monitoring
- Subthalamic Nucleus