Randomized Comparison of 3 High-Level Disinfection and Sterilization Procedures for Duodenoscopes.
rct · Level II
Where this comes from
- Record sourced from PubMed, PMID 28711629.
- Also identified by DOI 10.1053/j.gastro.2017.06.052.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Duodenoscopes have been implicated in the transmission of multidrug-resistant organisms (MDRO). We compared the frequency of duodenoscope contamination with MDRO or any other bacteria after disinfection or sterilization by 3 different methods. We performed a single-center prospective randomized study in which duodenoscopes were randomly reprocessed by standard high-level disinfection (sHLD), double high-level disinfection (dHLD), or standard high-level disinfection followed by ethylene oxide gas sterilization (HLD/ETO). Samples were collected from the elevator mechanism and working channel of each duodenoscope and cultured before use. The primary outcome was the proportion of duodenoscopes with an elevator mechanism or working channel culture showing 1 or more MDRO; secondary outcomes included the frequency of duodenoscope contamination with more than 0 and 10 or more colony-forming units (CFU) of aerobic bacterial growth on either sampling location. After 3 months of enrollment, the study was closed because of the futility; we did not observe sufficient events to evaluate the primary outcome. Among 541 duodenoscope culture events, 516 were included in the final analysis. No duodenoscope culture in any group was positive for MDRO. Bacterial growth of more than 0 CFU was noted in 16.1% duodenoscopes in the sHLD group, 16.0% in the dHLD group, and 22.5% in the HLD/ETO group (P = .21). Bacterial growth or 10 or more CFU was noted in 2.3% of duodenoscopes in the sHLD group, 4.1% in the dHLD group, and 4.2% in the HLD/ETO group (P = .36). MRDOs were cultured from 3.2% of pre-procedure rectal swabs and 2.5% of duodenal aspirates. In a comparison of duodenoscopes reprocessed by sHLD, dHLD, or HLD/ETO, we found no significant differences between groups for MDRO or bacteria contamination. Enhanced disinfection methods (dHLD or HLD/ETO) did not provide additional protection against contamination. However, insufficient events occurred to assess our primary study end-point. ClinicalTrials.gov no: NCT02611648.
Medical subject headings
- Cross Infection
- Disinfectants
- Disinfection
- Drug Resistance, Multiple, Bacterial
- Duodenoscopes
- Duodenoscopy
- Equipment Contamination
- Equipment Reuse
- Ethylene Oxide
- Sterilization
- o-Phthalaldehyde