A Parametric Computational Analysis Into Galvanic Coupling Intrabody Communication.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 28783652.
- Also identified by DOI 10.1109/JBHI.2017.2734939.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Intrabody Communication (IBC) uses the human body tissues as transmission media for electrical signals to interconnect personal health devices in wireless body area networks. The main goal of this work is to conduct a computational analysis covering some bioelectric issues that still have not been fully explained, such as the modeling of skin-electrode impedance, the differences associated with the use of constant voltage, or current excitation modes, or the influence on attenuation of the subject's anthropometrical and bioelectric properties. With this aim, a computational finite element model has been developed, allowing the IBC channel attenuation as well as the electric field and current density through arm tissues to be computed as a function of these parameters. As a conclusion, this parametric analysis has in turn permitted us to disclose some knowledge about the causes and effects of the above-mentioned issues, thus, explaining and complementing previous results reported in the literature.
Medical subject headings
- Electric Impedance
- Galvanic Skin Response
- Models, Biological
- Telemetry