Evaluation of the Boussignac Cardiac arrest device (B-card) during cardiopulmonary resuscitation in an animal model.

Moore, Johanna C; Lamhaut, Lionel; Hutin, Alice; Dodd, Kenneth W; Robinson, Aaron E; Lick, Michael C; Salverda, Bayert J; Hinke, Mason B et al. · Resuscitation · 2017

basic_science · Level V

Where this comes from

Abstract

The purpose of this study was to examine continuous oxygen insufflation (COI) in a swine model of cardiac arrest. The primary hypothesis was COI during standard CPR (S-CPR) should result in higher intrathoracic pressure (ITP) during chest compression and lower ITP during decompression versus S-CPR alone. These changes with COI were hypothesized to improve hemodynamics. The second hypothesis was that changes in ITP with S-CPR+COI would result in superior hemodynamics compared with active compression decompression (ACD) + impedance threshold device (ITD) CPR, as this method primarily lowers ITP during chest decompression. After 6min of untreated ventricular fibrillation, S-CPR was initiated in 8 female swine for 4min, then 3min of S-CPR+COI, then 3min of ACD+ITD CPR, then 3min of S-CPR+COI. ITP and hemodynamics were continuously monitored. During S-CPR+COI, ITP was always positive during the CPR compression and decompression phases. ITP compression values with S-CPR+COI versus S-CPR alone were 5.5±3 versus 0.2±2 (p<0.001) and decompression values were 2.8±2 versus -1.3±2 (p<0.001), respectively. With S-CPR+COI versus ACD+ITD the ITP compression values were 5.5±3 versus 1.5±2 (p<0.01) and decompression values were 2.8±2 versus -4.7±3 (p<0.001), respectively. COI during S-CPR created a continuous positive pressure in the airway during both the compression and decompression phase of CPR. At no point in time did COI generate a negative intrathoracic pressures during CPR in this swine model of cardiac arrest.

Medical subject headings