A Biomechanical Comparison Between Asymmetric Pennington Technique and Conventional Core Suture Techniques: 6-Strand Flexor Tendon Repair.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 28807348.
- Also identified by DOI 10.1016/j.jhsa.2017.06.104.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To evaluate the fatigue strength and gap sizes of the asymmetric Pennington technique compared with 2 conventional 6-strand core suture techniques: the triple-looped suture and the Yoshizu #1. We recorded the fatigue strength (forces × cycles) and gap sizes of a 6-strand flexor tendon repair with different core suture techniques under cyclic loading in 30 porcine tendons. The asymmetric Pennington technique was performed with a Pennington repair of equal suture purchase in the 2 tendon stumps, with the 2 other Pennington repairs shifted by 3 mm, respectively, along the longitudinal axis of the tendon in relation to the first Pennington repair. The triple-looped suture technique was made with triple Tsuge sutures. The Yoshizu #1 technique was performed with a combined Pennington repair (using a double strand) and Tsuge suture. The asymmetric Pennington technique showed significantly greater fatigue strength and significantly smaller gaps in comparison to the triple-looped suture and Yoshizu #1 techniques. This study demonstrated that the asymmetric Pennington technique generated increased fatigue strength and reduced gap sizes compared with 2 conventional 6-strand core suture techniques, the triple-looped suture and Yoshizu #1. The asymmetric Pennington technique may permit an early active motion rehabilitation protocol similar to the triple-looped suture and Yoshizu #1 techniques.
Medical subject headings
- Materials Testing
- Stress, Mechanical
- Suture Techniques
- Tendons