On-chip detection of gel transition temperature using a novel micro-thermomechanical method.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 28817711.
- Also identified by DOI 10.1371/journal.pone.0183492 and PMC identifier 5560686.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We present a new thermomechanical method and a platform to measure the phase transition temperature at microscale. A thin film metal sensor on a membrane simultaneously measures both temperature and mechanical strain of the sample during heating and cooling cycles. This thermomechanical principle of operation is described in detail. Physical hydrogel samples are prepared as a disc-shaped gels (200 μm thick and 1 mm diameter) and placed between an on-chip heater and sensor devices. The sol-gel transition temperature of gelatin solution at various concentrations, used as a model physical hydrogel, shows less than 3% deviation from in-depth rheological results. The developed thermomechanical methodology is promising for precise characterization of phase transition temperature of thermogels at microscale.
Medical subject headings
- Equipment Design
- Gels
- Lab-On-A-Chip Devices
- Temperature