Novel technique to accurately measure femoral diameter using a Thomas splint.
expert_opinion · Level V
Where this comes from
- Record sourced from PubMed, PMID 28916134.
- Also identified by DOI 10.1016/j.injury.2017.08.068.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
During surgical management of femoral shaft fractures, difficulties arise when treating patients with narrow femoral diaphyseal canals, such as young patients and those with dysplastic femurs secondary to underlying pathology. Accurate pre-operative assessment of the femoral diaphyseal canal diameter would allow the surgeon to plan surgical technique and ensure appropriate equipment was available, such as narrow, unreamed or paediatric sized nails. When secured to the patient both longitudinal rods of the main Thomas Splint component lie parallel with the femoral shaft and horizontal to the radiographic x-ray plate. The diameter of these rods are 13mm (Adult and paediatric). Using the calibration tool, we calibrate the diameter of the Thomas Splint to 13mm, accurately measuring any further detail on that radiograph, such as the diaphyseal canal diameter. Accurate knowledge pre-operatively of radiographic measurements is highly valuable to the operating surgeon. This technique can accurately measure femoral canal diameter using the Thomas splint, negates the requirement for a calibration marker, is reproducible, easy to perform, and is indispensible when faced with a patient with a narrow femoral canal in a diaphyseal femoral fracture. (181 words).
Medical subject headings
- Calibration
- Femoral Fractures
- Femur
- Radiography
- Splints