The influence of two common sterilization techniques on the corrosion of Mg and its alloys for biomedical applications.

Johnston, Sean; Shi, Zhiming; Hoe, Cherilyn; Uggowitzer, Peter J; Cihova, Martina; Löffler, Jörg F; Dargusch, Matthew S; Atrens, Andrej · J Biomed Mater Res B Appl Biomater · 2018

basic_science · Level V

Where this comes from

Abstract

This paper studied the influence of two common sterilization techniques, ethylene oxide (EO) and gamma irradiation (GI), on the corrosion rate of four Mg-based materials in CO<sub>2</sub> -bicarbonate buffered Hanks' solution. The four materials were: high-purity (HP)-Mg, ZE41, ultra-high purity (XHP)-Mg, and XHP-ZX00. The corrosion rate was measured through mass loss (P<sub>m</sub> ) and hydrogen evolution (P<sub>H</sub> ). Two-way analysis of variance (ANOVA) was conducted to assess the effect of the sterilization techniques on the corrosion rates across the four materials. The ANOVA analyzed the variables of (1) material, (2) sterilization condition (EO, GI, and an unsterilized control group), and (3) the interaction between these two independent variables. Neither sterilization technique (EO and GI) significantly influenced the corrosion rate as measured by P<sub>m</sub> (p < 0.84) nor P<sub>H</sub> (p < 0.08). This result was consistent across the four materials tested, as there was no interaction between the test variables of material and sterilization condition for P<sub>m</sub> (p < 0.49) or P<sub>H</sub> (p < 0.27). As neither EO nor GI influenced the corrosion rates, either of these techniques warrants consideration for use on Mg-based medical implants and devices. © 2017 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater, 106B: 1907-1917, 2018.