Bioabsorbable poly-L/D-lactide (96/4) scaffold arthroplasty (RegJoint™) for trapeziometacarpal osteoarthritis: a 3-year follow-up study.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 28975848.
- Also identified by DOI 10.1177/1753193417732002.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The poly-L/D-lactide joint scaffold (RegJoint™) has recently been associated with adverse tissue reactions and osteolysis after partial trapeziectomy for trapeziometacarpal osteoarthritis. Twenty-two of 23 patients previously operated on with this scaffold were re-examined at a mean follow-up of 3.3 years (range 36-53 months). Overall, the results showed an unacceptably high rate of adverse tissue reactions related to the degradation process of the implant, resulting in a revision procedure in three patients. At final follow-up, at which point the implant had completely degraded, there were no signs of ongoing adverse tissue reactions. There was a significant decrease in pain, increase in strength and subjective improvement in function at final follow-up compared with the pre-operative results in patients who had not undergone revision surgery. However, owing to the high incidence of adverse tissue reactions, the use of the implant has been discontinued in the treatment of trapeziometacarpal osteoarthritis. IV.
Medical subject headings
- Absorbable Implants
- Arthroplasty, Replacement
- Carpometacarpal Joints
- Joint Prosthesis
- Osteoarthritis
- Thumb
- Trapezium Bone
Anatomy
- hand
- wrist