Ultrastructural Integrity of Human Capsulotomies Created by a Thermal Device.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 29074029.
- Also identified by DOI 10.1016/j.ophtha.2017.09.018.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To investigate the ultrastructural features of anterior capsulotomy performed with a thermal device, the precision pulse capsulotomy (PPC). Prospective, multicenter case series. Consecutive patients undergoing capsulotomy with the PPC device. Prospective study of patients undergoing capsulotomy with the PPC by 2 surgeons, followed up by routine phacoemulsification cataract surgery, was undertaken. All capsulotomy specimens were collected for scanning electron microscopy (SEM). Observations were made regarding uniformity of the capsular edge and the presence of irregularities that may compromise integrity. Comparisons were made with manual continuous curvilinear capsulorrhexis. Ultrastructural features of PPC and presence of irregularities. Frayed appearance of the anterior capsule edge was noted in postoperative visits under slit-lamp examination. Scanning electron microscopy sampling showed a generally uniform rolled capsular edge, but interspersed with areas of irregularity with frayed appearance at the capsule margin. The PPC device is capable of creating reproducible, central, and precise circular capsulotomy. The ultrastructural features in ex vivo human capsulotomy specimens generally show eversion of the capsulotomy edge, but in some cases, this was accompanied by areas of irregular capsule margin with frayed edges, likely caused by dissipated thermal energy. The postoperative appearance and SEM features warrant further assessment of the PPC integrity and clinical correlation.
Medical subject headings
- Anterior Capsule of the Lens
- Capsulorhexis
- Laser Therapy
- Lasers