Effect of Beam Steering on Echogenic and Nonechogenic Needle Visibility at 40°, 50°, and 60° Needle Insertion Angles.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 29116966.
- Also identified by DOI 10.1213/ANE.0000000000002618.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Currently, there is little understanding of the role of echogenic needles and beam steering at moderate angles of needle insertion. The ultrasound images of the echogenic and nonechogenic needles inserted into pork at 40°, 50°, and 60° were scored by anesthesiologists on a scale of 0-10. The effect of different levels of beam steer was also explored. At 40°, steep beam steering improves visualization of both nonechogenic and echogenic needles to an equal, satisfactory level. At 50° and 60°, visualization of nonechogenic needles is poor, whereas visibility of an echogenic needle was adequate and may be improved with steep beam steering.
Medical subject headings
- Anesthesia, Conduction
- Needles
- Nerve Block
- Ultrasonography, Interventional