Enterprise Deployment Through PulseRider To Treat Anterior Communicating Artery Aneurysm Recurrence.

Valente, Iacopo; Limbucci, Nicola; Nappini, Sergio; Rosi, Andrea; Laiso, Antonio; Mangiafico, Salvatore · World Neurosurg · 2018

case_report · Level V

Where this comes from

Abstract

PulseRider (Pulsar Vascular, Los Gatos, California, USA) is a new endovascular device designed to treat wide-neck bifurcation intracranial aneurysms. Deployment of a stent through a PulseRider to treat an aneurysm's recurrence has never been described before. We report the case of a 55-year-old man who underwent coiling of an 8-mm anterior communicating artery aneurysm with assistance of a PulseRider neck reconstruction device. The 6-month digital subtraction angiography control showed aneurysm recurrence, so we deployed an Enterprise 2 closed-cell stent (Codman, Miami Lakes, Florida, USA) in the A1-A2 segment passing across the previously implanted PulseRider. Enterprise correctly expanded and allowed for adequate coiling of the aneurysm. An Enterprise stent can be safely opened through a PulseRider in order to treat aneurysm recurrence.