Wear rate and medium-term survival of a cemented, moderately cross-linked polyethylene acetabular prosthesis.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 29192734.
- Also identified by DOI 10.5301/hipint.5000521.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The moderately cross-linked Depuy Marathon® cemented acetabular component was introduced into the UK in 2007. The wear rate for the previously introduced Marathon® uncemented acetabular component has been reported to range from 0.06 to 0.01 mm/year. The aim of this study was to present the medium-term results and wear rate of the Marathon® cemented prosthesis used in primary total hip arthroplasty. 103 Marathon® cemented acetabular components were implanted between 2008 and 2009 in primary arthroplasty, who were eligible for this study. All patients received a metal 28-mm head. Mean age was 68 years (range 27-87). Mean clinical follow-up was 55 months (range 50-61). Mean radiological follow-up was 46 months (range 24-57). Wear was calculated on AP radiographs using computer-assisted uni-radiographic technique. The mean wear was 0.37 mm (range 0.0-0.78 mm). The wear rate was calculated as 0.03 mm/year (95% confidence interval 0.02-0.06). Postoperative complications included deep vein thrombosis (2%) and dislocation (0.8%); there were no deep infections. There were no revisions for failure of the Marathon® cemented acetabular component. The Marathon® cemented acetabular component demonstrates satisfactory wear rates and survivorship at medium-term follow-up.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Hip Joint
- Hip Prosthesis
- Joint Diseases
- Prosthesis Failure
Anatomy
- hip