Solvent or thermal extraction of ethylene oxide from polymeric materials: Medical device considerations.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 29226556.
- Also identified by DOI 10.1002/jbm.b.34052.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Ethylene oxide (EO) gas is commonly used to sterilize medical devices. Bioavailable residual EO, however, presents a significant toxicity risk to patients. Residual EO is assessed using international standards describing extraction conditions for different medical device applications. We examine a series of polymers and explore different extraction conditions to determine residual EO. Materials were sterilized with EO and exhaustively extracted in water, in one of three organic solvents, or in air using thermal desorption. The EO exhaustively extracted varies significantly and is dictated by two factors: the EO that permeates the material during sterilization; and the effectiveness of the extraction protocol in flushing residual EO from the material. Extracted EO is maximized by a close matches between Hildebrand solubility parameters δ<sub>polymer</sub> , δ<sub>EO</sub> , and δ<sub>solvent</sub> . There remain complexities to resolve, however, because maximized EO uptake and detection are accompanied by great variability. These observations may inform protocols for material selection, sterilization, and EO extraction. © 2017 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater, 106B: 2455-2463, 2018.
Medical subject headings
- Equipment and Supplies
- Ethylene Oxide
- Sterilization