Reducing fretting-initiated crevice corrosion in hip simulator tests using a zirconia-toughened alumina femoral head.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 29266687.
- Also identified by DOI 10.1002/jbm.b.34062.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Taper fretting corrosion is considered a potentially limiting factor for total hip arthroplasty longevity. Recently, attention has been focused on new materials for ceramic femoral heads, for example, zirconia-toughened alumina (ZTA), since they have an alternative bearing surface that can improve the wear resistance. Moreover, ceramics have high chemical stability and corrosion resistance. In this study, we evaluated the effects of ZTA and Co-Cr-Mo alloy femoral heads on their taper fretting and/or corrosion characteristics under a controlled hip simulator test. After the test, less fretting and corrosion were observed in the taper surface of the trunnion against the ZTA femoral head than for that against the Co-Cr-Mo alloy femoral head. In addition, corrosion damages were only observed in the lateral-distal taper surface (noncontact area) of the trunnion in the Co-Cr-Mo alloy femoral head group. The ZTA femoral head group also eliminated the potential for Co ion release into the lubricants from taper corrosion, reducing the possibility of adverse local tissue inflammatory responses. In conclusion, ZTA femoral heads showed markedly less fretting corrosion compared to Co-Cr-Mo alloy femoral heads and have a lower potential for metal ion release. © 2017 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater, 106B: 2815-2826, 2018.
Medical subject headings
- Aluminum Oxide
- Femur Head
- Hip Prosthesis
- Prosthesis Design
- Vitallium
- Zirconium