Detecting and managing device leads inadvertently placed in the left ventricle.
Where this comes from
- Record sourced from PubMed, PMID 29328900.
- Also identified by DOI 10.3949/ccjm.85a.17012.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Inadvertent malpositioning of a cardiac implantable electronic device lead into the left ventricle is a rare complication of transvenous pacing and defibrillation. Rapid identification of lead position is critical during implantation and just after the procedure, with immediate correction required if malpositioning is detected. If lead misplacement is discovered late after implantation, the lead should be surgically removed or chronic anticoagulation with warfarin should be instituted.
Medical subject headings
- Blood Vessel Prosthesis Implantation
- Defibrillators, Implantable
- Prosthesis Failure
- Ventricular Dysfunction, Left