How reliable are ADC measurements? A phantom and clinical study of cervical lymph nodes.

Moreau, Bastien; Iannessi, Antoine; Hoog, Christopher; Beaumont, Hubert · Eur Radiol · 2018

prospective_cohort · Level II

Where this comes from

Abstract

To assess the reliability of ADC measurements in vitro and in cervical lymph nodes of healthy volunteers. We used a GE 1.5 T MRI scanner and a first ice-water phantom according to recommendations released by the Quantitative Imaging Biomarker Alliance (QIBA) for assessing ADC against reference values. We analysed the target size effect by using a second phantom made of six inserted spheres with diameters ranging from 10 to 37 mm. Thirteen healthy volunteers were also scanned to assess the inter- and intra-observer reproducibility of volumetric ADC measurements of cervical lymph nodes. On the ice-water phantom, the error in ADC measurements was less than 4.3 %. The spatial bias due to the non-linearity of gradient fields was found to be 24 % at 8 cm from the isocentre. ADC measure reliability decreased when addressing small targets due to partial volume effects (up to 12.8 %). The mean ADC value of cervical lymph nodes was 0.87.10<sup>-3</sup> ± 0.12.10<sup>-3</sup> mm<sup>2</sup>/s with a good intra-observer reliability. Inter-observer reproducibility featured a bias of -5.5 % due to segmentation issues. ADC is a potentially important imaging biomarker in oncology; however, variability issues preclude its broader adoption. Reliable use of ADC requires technical advances and systematic quality control. • ADC is a promising quantitative imaging biomarker. • ADC has a fair inter-reader variability and good intra-reader variability. • Partial volume effect, post-processing software and non-linearity of scanners are limiting factors. • No threshold values for detecting cervical lymph node malignancy can be drawn.

Medical subject headings