Intraoperative Comparison of Measured Resection and Gap Balancing Using a Force Sensor: A Prospective, Randomized Controlled Trial.
rct · Level II
Where this comes from
- Record sourced from PubMed, PMID 29602537.
- Also identified by DOI 10.1016/j.arth.2018.02.044.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
For establishing femoral component position, gap-balancing (GB) and measured resection (MR) techniques were compared using a force sensor. Ninety-one patients were randomized to undergo primary total knee arthroplasty using either MR (n = 43) or GB (n = 48) technique using a single total knee arthroplasty design. GB was performed with an instrumented tensioner. Force sensor data were obtained before the final implantation. GB resulted in greater range of femoral component rotation vs MR (1.5° ± 2.9° vs 3.1° ± 0.5°, P < .05) and posterior condylar cut thickness medially (10.2 ± 2.0 mm vs 9.0 ± 1.3 mm) and laterally (8.5 ± 1.9 mm vs 6.4 ± 1.0 mm). Force sensor data showed a decreased intercompartmental force difference at full flexion in GB (.8 ± 2.3 vs 2.0 ± 3.3u, 1u ≈ 15 N, P < .05). GB resulted in a greater range of femoral component rotation and thicker posterior condylar cuts resulting in an increased flexion space relative to MR. Intercompartmental force difference trended toward a more uniform distribution between full extension and full flexion in the GB vs MR group.
Medical subject headings
- Arthroplasty, Replacement, Knee
- Knee Joint
- Orthopedics
- Osteoarthritis, Knee
Anatomy
- femur
- knee