True rotational stability of a single-piece hydrophobic intraocular lens.
Where this comes from
- Record sourced from PubMed, PMID 29666120.
- Also identified by DOI 10.1136/bjophthalmol-2017-311797.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To evaluate rotation and its influencing factors of an aspheric one-piece hydrophobic acrylic intraocular lens (IOL) Vivinex XY1 during 6 months after operation. In this institutional trial, 122 eyes of 66 patients were implanted with a non-toric aspheric IOL Vivinex XY1 (Hoya Corporation, Tokyo, Japan). IOL alignment was assessed at the end of surgery, 1 hour, 1 week, 1 month and 6 months after implantation. Confounding factors such as axial length, presence of anterior fibrosis and randomised implantation in four different intended axes (0°, 45°, 90°, 135°) were evaluated. Decentration and tilt were measured using a Purkinje metre. Assessment of rotational stability was possible for 103 of 122 implanted IOLs 6 months after eye surgery. The median absolute rotation was 1.1° (range: 0°-5°). Rotation was significantly increased within the first hour after operation compared with later time-points (p<0.001). No correlation was found with axial length and rotation (Spearman's r=0.048, p=0.63). No significant difference was observed regarding different implantation axes (p=0.75). Rotation was not influenced by the presence of anterior fibrosis (p=0.98). Assessing the true IOL position at the end of surgery is crucial for the evaluation of rotational stability of IOLs. No IOL rotation exceeding 5° could be detected 6 months after surgery.
Medical subject headings
- Artificial Lens Implant Migration
- Lenses, Intraocular
- Phacoemulsification
- Rotation