Performance and comparability of laboratory methods for measuring ferritin concentrations in human serum or plasma: A systematic review and meta-analysis.
meta_analysis · Level I
Where this comes from
- Record sourced from PubMed, PMID 29723227.
- Also identified by DOI 10.1371/journal.pone.0196576 and PMC identifier 5933730.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Different laboratory methods are used to quantify ferritin concentrations as a marker of iron status. A systematic review was undertaken to assess the accuracy and comparability of the most used methods for ferritin detection. National and regional databases were searched for prospective, retrospective, sectional, longitudinal and case-control studies containing the characteristics and performance of at least one method for serum/plasma ferritin determinations in humans published to date. The analysis included the comparison between at least 2 methods detailing: sensitivity, precision, accuracy, predictive values, inter-methods adjustment, and use of international reference materials. Pooled method performance was analyzed for each method and across methods. Search strategy identified 11893 records. After de-duplication and screening 252 studies were assessed, including 187 studies in the qualitative analysis and 148 in the meta-analysis. The most used methods included radiometric, nonradiometric and agglutination assays. The overall within-run imprecision for the most reported ferritin methods was 6.2±3.4% (CI 5.69-6.70%; n = 171), between-run imprecision 8.9±8.7% (CI 7.44-10.35%; n = 136), and recovery rate 95.6% (CI 91.5-99.7%; n = 94). The pooled regression coefficient was 0.985 among all methods analyzed, and 0.984 when comparing nonradiometric and radiometric methods, without statistical differences in ferritin concentration ranging from 2.3 to 1454 μμg/L. The laboratory methods most used to determine ferritin concentrations have comparable accuracy and performance. Registered in PROSPERO CRD42016036222.
Medical subject headings
- Blood Chemical Analysis
- Ferritins