Clinical measurement of the dart throwing motion of the wrist: variability, accuracy and correction.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 29754522.
- Also identified by DOI 10.1177/1753193418773329 and PMC identifier 6104201.
- Licence recorded as CC BY.
- The licence permits redistribution, so the abstract is shown in full and the full text is available from the publisher.
Abstract
Despite being functionally important, the dart throwing motion is difficult to assess accurately through goniometry. The objectives of this study were to describe a method for reliably quantifying the dart throwing motion using goniometric measurements within a healthy population. Wrist kinematics of 24 healthy participants were assessed using goniometry and optical motion tracking. Three wrist angles were measured at the starting and ending points of the motion: flexion-extension, radial-ulnar deviation and dart throwing motion angle. The orientation of the dart throwing motion plane relative to the flexion-extension axis ranged between 28° and 57° among the tested population. Plane orientations derived from optical motion capture differed from those calculated through goniometry by 25°. An equation to correct the estimation of the plane from goniometry measurements was derived. This was applied and differences in the orientation of the plane were reduced to non-significant levels, enabling the dart throwing motion to be measured using goniometry alone.
Medical subject headings
- Range of Motion, Articular
- Wrist Joint
Anatomy
- wrist