Sites of fractures in explanted NeuFlex® silicone metacarpophalangeal joint prostheses.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 29754523.
- Also identified by DOI 10.1177/1753193418774500.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Single-piece silicone implants dominate metacarpophalangeal joint arthroplasty. The NeuFlex® implant was introduced to improve on the clinical performance of other silicone implants by having a pre-flexed hinge. By visually examining a cohort of 30 explanted NeuFlex® metacarpophalangeal joint prostheses we sought to identify the failure modes of these implants. Seven were not fractured, 11 had fractured across the hinge, nine had fractured at the junction of the distal stem and the hinge, and three showed fractures at both the hinge and at the junction of the distal stem and the hinge. These data may prove helpful in identifying how the performance of single-piece silicone implant designs can be improved.
Medical subject headings
- Arthroplasty, Replacement, Finger
- Joint Prosthesis
- Metacarpophalangeal Joint
- Prosthesis Failure
Anatomy
- hand