Routine single particle CryoEM sample and grid characterization by tomography.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 29809143.
- Also identified by DOI 10.7554/eLife.34257 and PMC identifier 5999397.
- Licence recorded as CC0.
- The licence permits redistribution, so the abstract is shown in full and the full text is available from the publisher.
Abstract
Single particle cryo-electron microscopy (cryoEM) is often performed under the assumption that particles are not adsorbed to the air-water interfaces and in thin, vitreous ice. In this study, we performed fiducial-less tomography on over 50 different cryoEM grid/sample preparations to determine the particle distribution within the ice and the overall geometry of the ice in grid holes. Surprisingly, by studying particles in holes in 3D from over 1000 tomograms, we have determined that the vast majority of particles (approximately 90%) are adsorbed to an air-water interface. The implications of this observation are wide-ranging, with potential ramifications regarding protein denaturation, conformational change, and preferred orientation. We also show that fiducial-less cryo-electron tomography on single particle grids may be used to determine ice thickness, optimal single particle collection areas and strategies, particle heterogeneity, and de novo models for template picking and single particle alignment.
Medical subject headings
- Cryoelectron Microscopy
- Electron Microscope Tomography