Relationship between effective lens position and axial position of a thick intraocular lens.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 29902254.
- Also identified by DOI 10.1371/journal.pone.0198824 and PMC identifier 6001946.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To discuss the impact of intraocular lens-(IOL)-power, IOL-thickness, IOL-shape, corneal power and effective lens position (ELP) on the distance between the anterior IOL vertex (ALP) of a thick IOL and the ELP of its thin lens equivalent. We calculated the ALP of a thick IOL in a model eye, which results in the same focal plane as a thin IOL placed at the ELP using paraxial approximation. The model eye included IOL-power (P), ELP, IOL-thickness (Th), IOL-shape-factor (X), and corneal power (DC). The initial values were P = 10 D (diopter: 1 D = 1 m-1), 20 D, 30 D, Th = 0.9 mm, ELP = 5 mm, X = 0, DC = 43 D. The difference between ALP and the ELP was illustrated as a function of each of the model parameters. The ALP of a thick lens has to be placed in front of the ELP for P>0 IOLs to achieve the same optical effect as the thin lens equivalent. The difference ALP-ELP for the initial values is -0.57 mm. Minus power IOLs (ALP-ELP = -0.07 mm, for IOL-power = -5 D) and convex-concave IOLs (ALP-ELP = -0.16 mm, for X = 1) have to be placed further posterior. The corneal power and ELP have less influence, but corneal power cannot be neglected. The distance between ELP and ALP primarily depends on IOL-power, IOL-thickness, and shape-factor.
Medical subject headings
- Lenses, Intraocular
- Models, Biological