An Ultrasensitive Diagnostic Biochip Based on Biomimetic Periodic Nanostructure-Assisted Rolling Circle Amplification.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 29924598.
- Also identified by DOI 10.1021/acsnano.8b01950.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Developing portable and sensitive devices for point of care detection of low abundance bioactive molecules is highly valuable in early diagnosis of disease. Herein, an ultrasensitive photonic crystals-assisted rolling circle amplification (PCs-RCA) biochip was constructed and further applied to circulating microRNAs (miRNAs) detection in serum. The biochip integrated the optical signal enhancement capability of biomimetic PCs surface with the thousand-fold signal amplification feature of RCA. The biomimetic PCs displayed periodic dielectric nanostructure and significantly enhanced the signal intensity of RCA reaction, leading to efficient improvement of detection sensitivity. A limit of detection (LOD) as low as 0.7 aM was obtained on the PCs-RCA biochip, and the LOD was 7 orders of magnitude lower than that of standard RCA. Moreover, the PCs-RCA biochip could discriminate a single base variation in miRNAs. Accurate quantification of ultralow-abundance circulating miRNAs in clinical serum samples was further achieved with the PCs-RCA biochip, and patients with the nonsmall cell lung carcinoma were successfully distinguished from healthy donors. The PCs-RCA biochip can detect bioactive molecules with ultrahigh sensitivity and good specificity, making it valuable in clinical disease diagnosis and health assessment.
Medical subject headings
- Carcinoma, Non-Small-Cell Lung
- Lung Neoplasms
- MicroRNAs
- Nanostructures
- Nucleic Acid Amplification Techniques