Methodological Shortcomings of Wrist-Worn Heart Rate Monitors Validations.
Where this comes from
- Record sourced from PubMed, PMID 29967000.
- Also identified by DOI 10.2196/10108 and PMC identifier 6048383.
- Licence recorded as CC BY.
- The licence permits redistribution, so the abstract is shown in full and the full text is available from the publisher.
Abstract
Wearable sensor technology could have an important role for clinical research and in delivering health care. Accordingly, such technology should undergo rigorous evaluation prior to market launch, and its performance should be supported by evidence-based marketing claims. Many studies have been published attempting to validate wrist-worn photoplethysmography (PPG)-based heart rate monitoring devices, but their contrasting results question the utility of this technology. The reason why many validations did not provide conclusive evidence of the validity of wrist-worn PPG-based heart rate monitoring devices is mostly methodological. The validation strategy should consider the nature of data provided by both the investigational and reference devices. There should be uniformity in the statistical approach to the analyses employed in these validation studies. The investigators should test the technology in the population of interest and in a setting appropriate for intended use. Device industries and the scientific community require robust standards for the validation of new wearable sensor technology.
Medical subject headings
- Heart Rate
- Monitoring, Physiologic
- Wearable Electronic Devices
- Wrist