Design of a Quadrature 1H/31P Coil Using Bent Dipole Antenna and Four-Channel Loop at 3T MRI.

Hong, Suk-Min; Choi, Chang-Hoon; Magill, Arthur W; Jon Shah, N; Felder, Jarg · IEEE Trans Med Imaging · 2018

basic_science · Level V

Where this comes from

Abstract

MRI using nuclei other than protons is of clinical interest due to the important role of these nuclei in cellular processes. Phosphorous-31 (<sup>31</sup>P), for example, plays an important role in energy metabolism. However, measurement of <sup>31</sup>P can be challenging, as the receive signal is weak compared with that of proton (<sup>1</sup>H). Consequently, it is often necessary to integrate <sup>1</sup>H elements for localizations and B<sub>0</sub> shimming in RF coils intended for <sup>31</sup>P measurements. Good decoupling between the <sup>1</sup>H and the <sup>31</sup>P elements is therefore essential. In this paper, bent dipole antennas tuned to <sup>1</sup>H were integrated with a four channel <sup>31</sup>P loop coil array, in a manner providing strong geometric decoupling between dipoles and loops. As the physical length of a resonant dipole antenna is too long at 3T, the dipole antennas were bent around the load. The loss of <sup>31</sup>P elements due to the presence of the dipole antennas was evaluated by measuring scattering parameters and comparing the SNR of <sup>31</sup>P spectra with and without the presence of the dipole antennas. The performance of the bent dipole antenna was evaluated by simulation and sensitivity measurement. The Q-factors and the SNR of the four-loop array were reduced by less than 5% when the bent dipole antennas were introduced. The measured sensitivity of the bent dipole was higher (15%) than that of dual-tuned birdcage. The combined bent dipole and loop array is therefore a promising design for <sup>1</sup>H/<sup>31</sup>P applications at 3T.

Medical subject headings